TMS Europe LtdTMS Europe Ltd is a UKAS accredited calibration laboratory No.0461, as defined in our Schedule of Accreditation. As a UK manufacturer we're a member of Made in Britain.

Type N MI Thermocouples, Batch UKAS Calibrated for AMS 2750, 3mm diameter and 75mm tails

UKAS Calibrated
Type N MI Thermocouples, Batch UKAS Calibrated for AMS 2750, 3mm diameter and 75mm tails
ISO 17025 Calibrated
[ Thermocouple Diagram - Type N MI Thermocouples, Batch UKAS Calibrated for AMS 2750, 3mm diameter and 75mm tails]

TMS Stock Code:   EXAMSN3MMX#

Length

Available for ordering online (UK+IE) or off-line.
[UKAS 0461 Logo]

Includes UKAS Calibration (at TMS standard points)

TMS Europe is a UKAS (ISO 17025) accredited calibration laboratory No. 0461. We provide this item complete with a UKAS calibration certificate, with customer details, listing that unit's results at our stated standard points. For different calibration points, customer details, etc, please contact us with your requirements for a quote.
Batch Calibration results are of the two samples from each end of the material roll used to make the thermocouples under the AMS 2750 aerospace standard. (This is distinct from calibration of individual units, as may be required if working to other standards.)
All available options are shown below.
Length
Price (Ex VAT)Estimated
Availability
Estimated
Stock
2.5m
£52.541 - 4 days 0
3m
£63.041 - 4 days 0
4m
£81.861 - 4 days 0
5m
£91.261 - 4 days 0
6m
£99.861 - 4 days 0
12m
£181.021 - 4 days 0
2m
£45.721 - 4 days 0

Description and Specifications
Description and Specifications

Type N Mineral Insulated (MI) Inconel sheathed thermocouples in a 3mm diameter with 75mm PFA leads and terminating in a mini plug. These thermocouples are available in 1, 1.5, 2, 2.5, 3, 4, 5, 6 and 12 metre sheath lengths and are suitable for temperatures up to 1220°C. These thermocouples are produced from batch calibrated cable and conform to AMS 2750.

These thermocouples are AMS 2750 and Nadcap compliant and suitable for Secondary Standard, Temperature Uniformity Surveys, System Accuracy Tests (3.1.6), Control Recording and Monitoring and Load . Each thermocouple has a unique label attached, giving full traceability to a UKAS Calibration Certificate, which gives the calibration results of the two samples for each end of the roll used to make the thermocouples.

Temperature Ranges: +20°C to 1220°C (20 - 1100°C for continuous use)
Batch Calibration Points: 20, 100, 200, 300, 400, 500, 550, 600, 650, 700, 750, 800, 900, 1000, 1100, 1200 & 1220°C

One Calibration Certificate is supplied with every order.
Note: These temperatures apply to the MI sheath, the pot seal temperature and tails should not exceed 135°C.