Type N Mineral Insulated (MI) Inconel sheathed thermocouples in a 2mm diameter terminating in a fixed mini plug. Available in a 3.5 metre sheath length and are suitable for temperatures up to 1220°C, produced from batch calibrated cable they conform to AMS 2750. Supplied with a UKAS calibration certificate.
These thermocouples are AMS 2750 and Nadcap compliant and suitable for Secondary Standard, Temperature Uniformity Surveys, System Accuracy Tests (3.1.6), Control Recording and Monitoring and Load . Each thermocouple has a unique label attached, giving full traceability to a UKAS Calibration Certificate, which gives the calibration results of the two samples for each end of the roll used to make the thermocouples.
Temperature Ranges: +20°C to 1220°C (20 - 1100°C for continuous use)
Batch Calibration Points: 100, 200, 300, 400, 500, 600, 700, 800, 900, 1000, 1100, 1200 & 1250°C
One Calibration Certificate is supplied with every order.
Note: These temperatures apply to the MI sheath, the pot seal temperature and tails should not exceed 135°C.